About air ions:
Almost all "+" natural ions come from radioactivity. About
40% of natural air ions come from radioactive minerals in the ground.
Each time a radioactive atom decays near the air, it produces 50,000
- 500,000 air ion pairs. Another 40% comes from radon in the air
(which produces about 250,000 ion pairs for each radon atom), and
20% comes from cosmic rays (high-energy protons from distant supernovas).
Indoors, ions "live" on average about 30 seconds before
touching a surface and shorting to ground. Outdoor ions usually "live" several
minutes. Negative ions come from radioactivity and evaporating water.
Also lightning, thunderstorms, and forest fires contribute "+" and "-" ions,
but since these ions are not produced during fair weather, it is
usually only radioactivity and evaporating water that produce ions
outdoors. Normal fair-weather ion concentrations are 200 to 800 negative
and 250 to 1500 positive ions per cubic centimeter. Indoor levels
are usually lower. Several hours before a storm, + ion concentration
will increase dramatically, sometimes exceeding 5000 ions/cm3 During
a storm, - ions increase to several thousand while + ions decrease,
often to below 500.
Because a large concentration of + ions can attract - ions, high
concentrations of + and - ions are often found together. Typically,
a high concentration (1000 or more) of both may be found in one area
outdoors while low concentration (300 or less) is found typically
one city block away. A cloud of pure + ions (no -) with a concentration
of 1000 ions/cm3 would be very unstable and would fall
apart if its diameter were more than about 30m (100'). For this reason,
high concentrations of exclusively + (or exclusively -) ions tend
to be compact, and don't extend more than about 30 m. While testing
indoors, you may find high - in one area of a room and high + in
another.
The life time of "fast" ions (these are the most common
type) is determined by how long they last before they collide with
a solid (or dust) which usually neutralizes their charge. Indoors
electric fields are stronger than outdoors. Plastic surfaces charge
to a typical potential of negative 1000 volts. This produces electric
fields of 500-5000 volts per meter near the plastic surface. The
electric field repels negative ions (air molecules with an extra
0- or 0H-. The mobility of fast - ions is about 1.2x104 m/s
per V/m, so at 2000 V/m, - ions are repelled at a speed of 2000x
1.2x10~4 =.24m/s (meters per second). Positive ions (air
molecules with an extra H+ or positive ammonia molecule) are attracted to
the plastic by the same field. Their mobility is slightly lower (about
1.0 x 10-4 m/s per V/m) and they have a slightly slower speed of
.2 m/s. When the + ions touch the plastic, they give up their + charge.
This partially neutralizes the - charge on the plastic. Under typical
conditions, complete neutralization of the -charge on the plastic
would occur in a few weeks. However, dust blowing by will rub against
the plastic and acquire a + charge. This dust carries the + charge
away (ultimately to Earth ground). As a result, the plastic always
retains a negative charge.
A good way to standardize (and lengthen) the lifetime of indoor
ions is to put them in a large cardboard box. Lifetime then is about
50 sec, regardless of humidity, so if, for example, 4 pCi/L of radon
is in the box, it will produce a continuous 1600 + ions/cm3 in
the box.
You can produce negative ions directly by combing your hair with
a plastic comb. If you then blow air past the comb, the air will
have between 1000 and 10,000 - ions/cm3 immediately next
to the comb. The number is lower in high humidity. Also, your breath
contains about 20,000 to 50,000 - ions/cm3 from the evaporating
water, but you must be grounded to exhale a concentration this high.
If you are insulated from ground, you will become more positively
charged with each exhalation (by about 5 volts) because your breath
is removing negative charge. Eventually, you will become sufficiently
positive (after exhaling about 20 times), that the negative ions
will immediately return to you. This is the same effect that occurs
in building cooling systems that use an evaporating water tower.
If not properly grounded, the water pump and vents will become very
positive. (If the inside vents are isolated from the evaporating
water via a heat exchanger, the vents may become very positive and
produce a large number of + ions. This can be corrected simply by
grounding the vent).
Both + and - ions come from combustion (flame, wood burning, cigarette
smoke, and car exhaust) and from very hot surfaces (hot enough to
glow).
Indoors, near ground level (basement), most "+" ions come
from radon, and a reading of 1000 "+" ions/cm3 means
about 4 pCi/L of radon; the maximum allowable amount in the U.S.
(This number of ions is directly proportional to radon concentration multiplied
by average ion lifetime:
strong electric fields indoors will reduce the ion lifetime.) Because
it is unlikely that a level so high (1000, or 1.00 on the counter)
can come from anything else (other than flame, smoke, or a hot electric
heating element), it is likely that 1000 ions/cm3 in a
basement means about 4 pCi/L of radon are present (or 2000 ions/cm3 =
8 pCi/L, etc.). Note that if radon is the source of the ions, then
the concentration of ions will be approximately equal throughout
the basement. If it is instead 1000 near a hot water heater but only
100 ions/cm3 elsewhere, it is not radon. A higher
concentration of + ions near cracks in the concrete foundation or
near corners indicate the radon is coming in there.
If the average "+" ion count is low (for example, less
than 100), then there is essentially no radon present. It is not
possible to "hide" the ions that radon produces. "No
ions" means "no radon". Occasionally, a small piece
of dust will discharge on the plate. Dust is usually "-",
so the plate will read typically -1000 (-1.00) or so, even if the
POLARITY switch is on "+". This will return to normal in
about 3 seconds. Holding the counter near any alpha particle source
(Uranium, Thorium, etc.) will produce very high ion readings, especially "+".
This Ion Counter can therefore directly be used in place of a Geiger
Counter. Turn the switch to STANDBY and polarity to "+".
Remove the wind guard and hold the top of the Ion Counter close to
the test source of possible radioactivity. If .1 microCurie of 5
to 8 MeV alpha (Uranium, Thorium, Radium) is entering the top hole,
the display will read 250,000 ions/cm3 ("250" on
the 1999 scale). These alpha particles can only travel through about
5 cm (2") of air, so hold the top of the counter very near the
suspected source. The display is proportional to the radioactivity
present. Neutrons can also be detected by putting a thin layer of
plastic (a hydrogen source) over the rectangular slot. This will
convert hi-energy neutrons to protons, which can be detected because
protons create ion pairs. Sensitivity is a few orders of magnitude
less than sensitivity to alpha particles (as described above).
INSTRUCTIONS
1. Start with the upper right switch on STANDBY
and the knob at OFF. Then turn the knob to 19.99. This will start
a 50 second warm-up. During this warm-up, all 3 decimal points will
be black. After warm-up, only one decimal point will remain. Then
push the right switch down to RE-ZERO and hold there at least 5 seconds.
The display should now read between 0.02 (positive) and -0.02 (negative)
and it should be stable. That is, it should stay in that range at
least 5 seconds. If the display is not stable, RE-ZERO again. Then
check that the wind is not blowing. If air is being forced through
the Ion Counter, shield the front with your hand and the back with
your body and wait until the display becomes stable. Then RE-ZERO
again. Alternatively, in windy conditions, you can balance the left
switch (POLARITY) in the center position (not + or-) and wait
for the display to become stable. Then RE-ZERO. This method will
produce a more accurate "zero" in turbulent conditions,
but you will also have to wait longer before taking measurements,
because you will need to switch away from this neutral position to
either + or -, which is an additional step.
2. To take a measurement, make sure POLARITY is
set either + or - (A change in the POLARITY setting will cause a
short delay. See 4.), and hold the Ion Counter away from you (at
arm's length) or set it down. This is because the synthetic fibers
in clothing often repel ions. Turn the switch to MEASURE, which will
turn on the fan and make all 3 decimal points visible for about 20
seconds. After 20 seconds, only the center decimal point will remain.
The ion counter works best if connected to ground or if it is momentarily
touched to a grounded object before measurement starts. To do this,
touch your finger to the center screw in a wall switch plate or wall
outlet while holding the counter, or touch metal plumbing or the
ground itself. If the ion counter has not been grounded, then it
may carry a significant static charge. This charge will either repel
or attract ions, which will distort the readings. Note that all of
the black surface of the ion counter (including the 4 rubber feet)
is electrically conductive.
3. Now the Ion Counter is ready to take measurements
and will display in units of 1000 ionslcm3, so multiply
the display by 1000. For example, -0.28 = 280 negative ions/cm3.
The air is sampled at the slot in the top of the Ion Counter. Hold
the Ion Counter in one place at least 10 seconds on MEASURE until
the reading is stable to get an accurate reading. However,
to determine a trend (whether the number of ions is greater
or less as you move to a new area), simply watch the display for
increase or decrease as you move the counter through various areas.
An ion "hot spot" or a depleted area will be apparent almost
immediately, even if you don't wait 10 seconds at that spot.
4. You can switch the POLARITY switch at any time.
The 3 decimal points will go black for approximately 30 seconds and
the fan will turn off, then back on by itself. You do not need to
RE-ZERO. Note that when the fan turns itself on, the 3 decimal points
will remain for an additional 20 seconds, as in step 2.
5. Most readings can be done in the most sensitive
range (19.99). If the scale indicates over range (1... or -1...),
switch to 199.9 or 1999. You must RE-ZERO after you switch range
(this will be obvious). To RE-ZERO, first switch to STANDBY and wait
about 20 seconds for the display to become stable.
6. RE-ZERO every 5 minutes (or more often if the
temperature changes rapidly). To do this, switch to STANDBY and wait
about 20 seconds until the display stops changing. If a breeze is
blowing, shield the Ion Counter during this 20 seconds, or switch
POLARITY to neutral as in 1. above. Then once the display becomes
stable (that is, it doesn't change by more than + .01 in 5 seconds)
hold the switch at RE-ZERO at least 3 seconds.
7. When about 15 minutes of battery life remains, "LO
BAT" indicator will come on at the left side of the liquid crystal
display, most likely when the switch is on MEASURE (fan "on").
This "LO BAT" (and the colon symbol, which will also show)
will fade out after about one minute, but as long as the batteries
are low, it will display again every time the right switch is changed
from STANDBY to MEASURE, or back. The Counter requires two 9-Volt
batteries. After heavy continuous use, LO BAT may come on prematurely.
Then just leave OFF a few hours so the batteries can rest. This will
prolong the battery life.
To replace batteries, unscrew the back (4 screws).
NOTES:
A. For fastest, most accurate readings, measure
an entire area on a single polarity, then reverse the polarity and
re-measure the entire area. (This is as opposed to measuring "+" then
immediately switching to "-" at each point in a room).
B. If ion concentration is less than 100 ("0.10")
a longer STANDBY settling time should be used before RE-ZERO. In
this case, watch the display until it stops drifting (about 30 seconds),
and then RE-ZERO. Also, the display will take longer to reach final
value when on MEASURE if the ion concentration is very low.
C. If you are measuring while walking, hold the
Ion Counter vertical and at arm's length. This will make the moving
air perpendicular to the air flow through the Ion Counter. In windy
conditions, hold it (or set it down) so it is vertical or at least
perpendicular to the wind direction. If changing conditions require
that it is sometimes parallel to the wind direction, have the air
flow in to the top (as opposed to into the bottom,
or fan side). This procedure will assure the most accurate readings.
When air is rapidly flowing into the top, the Ion Counter will read
slightly high, but if air is rapidly flowing into the fan side, it
may read very low.
D. While walking (especially on a carpet), or if
you are exposed to the output of an ionizer, you are acquiring charge.
This may attract or repel ions. The black outer casing of the Ion
Counter is conductive and will be at the same voltage as you are
while you are holding it. It too may attract or repel ions. When
walking on a carpet, best accuracy is obtained if you do not wear
shoes (or if the shoes are not plastic). Shoes which do not cause
you to be "shocked" when you touch a grounded object are
acceptable. Alternatively, while measuring, touch a grounded object
frequently (see 2.) to avoid too much buildup of charge. Also, you
can set the Ion Counter (on its back or standing) on a glass tray
or on a sheet of glass (plastic will not work well because it acquires
a charge). Hold the glass (not the Ion Counter) while walking and
touch the Ion Counter to ground just before starting measurements.
E. If the Ion Counter is at the same temperature
as the surroundings, the "ZERO" will be very stable. To
reach temperature equilibrium, leave the Ion Counter OFF at least
30 minutes in the environment that will be measured. If outside,
minimize its exposure to direct sunlight (which will make it hotter
than the environment).
F. MEASURE requires about five times the battery
power that STANDBY requires because MEASURE turns on the fan, so
keep on STANDBY when not taking measurements. Also, the fan is inhibited
from running during warm-up and polarity change, even when the switch
is on MEASURE. When turning OFF, the right switch can be either in
MEASURE or STANDBY.
G. To take the most accurate readings, make sure first that
the display reads near zero and remains near zero in STANDBY. If
blowing air or wind is moving through the ion counter while on standby,
the sensor will collect ions, thus producing a non-zero signal. In
that case, if you push down the switch to RE-ZERO, the new "zero" will
not be accurate. This is why, in instruction 1., you should shield
the ion counter when it's windy (or better yet, switch POLARITY to
center position) to obtain an accurate "zero".
H. Hold the counter at arm's length away from you
if you are wearing synthetic fiber clothing. This type of clothing
becomes charged and upsets the ion count. Also, the outside of the
case is coated with conductive paint. For most accurate readings,
you should touch your hand to a grounded object before switching
to MEASURE. This will "ground" the ion counter, so that
it will not repel ions. (If the outer case is charged "+" or "-" it
will either repel or absorb ions, so that either way, the counter
will not be accurate). Also, any strongly charged object will reduce
the apparent ion count if near the counter. In typical electric fields
near synthetic fabric, negative ions are repelled at a speed of around
3 to 30 cm/sec (about 1-10 inches/sec). Positive ions are similarly
attracted to the fabric.
I. Do not use solvents to clean the outside of
the ion counter. It may remove the special conductive paint. Use
only water.
J. The wind guard (black "handle" on
top) is conductive plastic. It is also an electrostatic shield, but
it can be removed to clean the metal plates inside. The plates only
need cleaning if a piece of dust or lint makes a bridge between the
center plate and the inside of the plate enclosure as shown.
K. Normally, when you reverse the POLARITY switch
while on STANDBY, the display should settle back down to near "zero'
after 30 seconds. If a dust bridge is present however, the display
will not settle near zero within 30 seconds after you reverse the
polarity switch. Instead, it will settle on +/- .10 or higher. For
example, if the display reads, "-0.01" on STANDBY, with
the polarity switch at "-", but after switching to "+" (and
then waiting 30 seconds on STANDBY), it reads "0.15", this
means that a piece of lint or other material (at least 5 mm long)
is forming a conductive path or bridge. To clean, snap out the wind
guard and blow air into the slot, or clear the space inside with
a strip of paper at least 10 cm (4") long to dislodge any material
that is forming a bridge in the gap. The counter will also work without
the wind guard in place, but the reading will be too high if wind
is blowing toward the top, and too low if wind is blowing in the
opposite direction. Also, without the wind guard (which is also a
conductive electrostatic shield), the counter could read too high
or too low while being moved toward or away from a highly charged
object. However, even if the wind guard is not in place, the counter
will read correctly after being held motionless about 20 seconds,
even if near a charged object.
L. When switching to MEASURE, the display may go
slightly negative, even when POLARITY is on or vice-versa. Then it
will recover. Within 20 seconds, this transient will settle and the
counter will read the correct ion count, and will continue to be
correct even as you move the counter through the area you're testing.
However, if you see a sudden increase or decrease in the ion count,
hold the counter still for at least 10 seconds for a correct new
reading. This will allow you to find areas with high (or unusually
low) ion concentration. Ion concentration may change significantly
depending on the time of the day and position (such as height above
the floor) inside a room.
M. The counter has an ion polarity selectivity
of approximately 20. This means, for example, that if there is a
very high positive ion concentration, the display may read positive
even when switched to negative. With this selectivity, 1000 positive
ions and no negative ions per cm3 will lead to a reading
of "1000" when POLARITY is on "+", and positive 50
(not negative) if set on "-". If the display shows that
one polarity of ions is more than five times as numerous as the other,
a more accurate reading will result if you subtract 1/20 of the high
number from the low. In the case above, 50 minus 1/20 of 1000 = "zero" negative
ions.
N. When the LO BAT display shows (when the batteries
become weak, it will first appear when on MEASURE), it means you
have approximately 15 minutes of MEASURE time left before the batteries
fail. (When they fail, the readings become unstable). Replace with
2 standard 9 Volt batteries ( alkalines will have about 3 times the
life of regular batteries). NiCad rechargeables work, but use the
8.4 Volt kind not 7.2 Volts. The colon symbol will also display
during low battery operation.
O. Ions are produced by high-energy events, such
as an open flame or a very hot object (hot enough to glow). Hot objects
usually emit equal numbers of "+" and "-'~ ions. In
addition, high DC voltage (over 1000 Volts), especially when connected
to pointed metal edges or needles, will produce ions of the same
polarity as the voltage source. This is the basis of home ionizers.
Evaporating water will produce "-" ions in the air and
as a consequence leave "+" charges behind in the water
that hasn't yet evaporated. If the excess "+" charges left
behind are not conducted back to ground, the water will become "+" enough
that "-" ion production will cease. For example, a fountain
that has a motor that plugs into the wall will continuously produce "-"ions
(until the water runs out) but a battery operated fountain will stop
producing "-' ions after a few minutes if the fountain is well
insulated from ground. The same is true of a battery-powered air
ionizer. In general, for every 3x10 13 water molecule that evaporates,
one water molecule carries an excess "-" charge.
P. You can set the polarity on "neutral" (neither "+" nor "-")
by balancing the POLARITY center position. Then the sign (+ or -)
of the display and its magnitude will tell you which type of ion
predominates in the air and its approximate concentration (although
the actual reading will only be about 1/10 the true sum of the +/-
ion concentrations). When switching back to "+" or "-" afterward,
allow about 30 seconds for the display to settle.
Q. It is possible that a small piece of lint or
string can interfere with the motion of the fan. Always check that
the fan comes on when it is supposed to, and if not, push it with
your finger, or blow on it to start. The fan is very low-force and
cannot hurt your finger. Check for and remove any material that may
interfere with the fan motion. The fan should operate whenever the
counter is on and the right switch is on MEASURE. The only two exceptions
are that the fan will not operate during the first 50 seconds after
the unit is turned on, or during the first 30 seconds after the POLARITY
switch is used.
SPECIFICATIONS:
The Air Ion Counter pulls air (or any other gas with ions present)
through a parallel plate assembly. Outer two plates are held at polarization
potential ("+" or "-"). Center is the linear
detector plate. Air gap is 4mm and polarization field is 1000 V/m.
Air flow: 200 cm3/sec (linear speed: 40 cm/sec)
Ion collection efficiency: 65%
Input resistance: 5 X 1010 Ohms
Dynamic range: 10 ions/cm3 (corresponds to 10 microVolts
at detector plate) to 2 million ions/cm3 (corresponds
to 2 Volts at detector plate)
Settling time: approximately 10 seconds
Noise level (10 seconds averaged): 10 ions/cm3
Accuracy: +/- 25% for fast ions (mobility greater than 8 x 15-5 m/s
per V/m - these are the most numerous ions. The Ion Counter is less
sensitive to "slow" ions such as charged pieces of dust).
Batteries: 2x 9V standard transistor radio or alkaline type.
Battery life: heavy duty 10 hours on STANDBY, or 2 hours on MEASURE.
Alkaline is 30 hours on STANDBY, or 6 hours on MEASURE. LO BAT indicator
comes on at 7.5 Volts (for each battery, or total 15 Volts). Indicates
10 to 30 minutes of MEASURE time remain.
Battery drain: 7 mA on STANDBY, 30 mA on MEASURE. Fails at 7 V/battery.
Ion Selectivity (crosstalk): 20x
Sensitivity with POLARITY switch in center position (neither "+" or "-")
: 1/10 of normal, but adds together "+" and "-"ions.
(Example: 600 "+" and 200 "-" will read 1/10
of 600 minus 200, or 40 ions/cm3).
Warranty: 1 year
Made in USA.
|